ADR-293: Vitals ground-truth rig — reference ingest, time alignment, and agreement metrics

  • Status: Accepted — initial implementation (this PR)
  • Date: 2026-08-10
  • Deciders: ruv
  • Tags: vitals, validation, ground-truth, bland-altman, evidence, honesty

Context

wifi-densepose-vitals (ADR-021) extracts breathing (0.10.5 Hz) and heart rate (0.82.0 Hz) from CSI. The 2026 research sweep found that every credible vitals result in the literature ships with reference-sensor ground truth (chest strap, pulse oximeter, ECG, or PSG), and that WiFi heart-rate numbers without stated scope (single person, static, line-of-sight, short range) are systematically misleading. RuView currently has no way to produce a MEASURED vitals number: there is no reference-signal ingest, no time alignment between CSI-derived estimates and a reference device, and no agreement statistics.

CLAUDE.md requires accuracy statements to be tagged MEASURED (with a reproducer), CLAIMED, or SYNTHETIC. For vitals, MEASURED is currently unreachable.

Options considered

  1. Live BLE/ANT+ integration with reference devices. Rejected for now: drivers and pairing are a hardware/product concern; the blocking gap is the evaluation math, not the radio link.
  2. File-based reference ingest + offline agreement analysis. Chosen: every consumer reference device (Polar, Garmin, oximeters) exports timestamped series; a file boundary keeps the crate dependency-free and the pipeline deterministic.

Decision

Add a groundtruth module to v2/crates/wifi-densepose-vitals:

1. Reference series ingest

  • ReferenceSeries: timestamped samples (unix millis + value) for one measurand (HeartRateBpm or BreathingRateBrpm), with device metadata (make/model, measurement principle). Parsed from CSV (timestamp_ms,value with a header line); malformed rows are rejected with row-numbered errors — untrusted file input validated at the boundary. Non-monotonic timestamps are an error, not silently sorted.

2. Time alignment

  • Constant-offset estimation by maximizing normalized cross-correlation of the estimate series against the reference over a bounded lag window (default ±30 s), on a common resampled grid (nearest-sample, no interpolation of physiological values across gaps larger than a configurable limit).
  • Optional linear clock-drift fit (offset + rate) for long sessions. Alignment parameters are reported, never silently applied.

3. Agreement metrics

  • AgreementReport: n paired samples, coverage fraction (time where both series had valid samples), MAE, RMSE, mean error (bias), BlandAltman 95% limits of agreement, and percentage-within-tolerance (configurable, default ±2 bpm HR / ±1 brpm breathing).
  • Session scope is mandatory metadata: subject count, motion state (static/moving), line-of-sight (LOS/NLOS/through-wall), distance band. A report without scope cannot be constructed.

4. Evidence tagging

  • EvidenceGrade::Measured is only constructible when the report carries a reference device, non-zero paired samples, minimum coverage, and a reproducer command string; otherwise the report grades as Claimed (real data, no reference) or Synthetic (generated input). This mirrors ADR-291's enforcement-in-types approach and the CLAUDE.md tagging rule.

Consequences

  • RuView can convert vitals claims from CLAIMED to MEASURED with a reproducible offline analysis, session by session, scope by scope.
  • Honest reporting will likely show heart-rate performance below marketing intuition, especially NLOS/moving — that is the purpose.
  • CSV ingest means a manual export step per session; acceptable at current scale, and the format is the de-facto export of consumer reference gear.
  • No clinical claim is implied: agreement statistics against consumer reference devices are engineering evidence, not medical validation.

Validation

  • cargo test -p wifi-densepose-vitals — CSV rejection cases, alignment recovery of known synthetic offsets/drifts, agreement metrics against hand-computed fixtures, evidence-grade constructibility rules.
  • cargo bench -p wifi-densepose-vitals — criterion benchmark for alignment over hour-scale synthetic sessions.
  • Real-session validation (ESP32 capture + chest strap) remains a follow-up requiring hardware evidence per CLAUDE.md.