# ADR-293: Vitals ground-truth rig — reference ingest, time alignment, and agreement metrics
|
||
|
||
- **Status**: Accepted — initial implementation (this PR)
|
||
- **Date**: 2026-08-10
|
||
- **Deciders**: ruv
|
||
- **Tags**: vitals, validation, ground-truth, bland-altman, evidence, honesty
|
||
|
||
## Context
|
||
|
||
`wifi-densepose-vitals` (ADR-021) extracts breathing (0.1–0.5 Hz) and heart
|
||
rate (0.8–2.0 Hz) from CSI. The 2026 research sweep found that every credible
|
||
vitals result in the literature ships with reference-sensor ground truth
|
||
(chest strap, pulse oximeter, ECG, or PSG), and that WiFi heart-rate numbers
|
||
without stated scope (single person, static, line-of-sight, short range) are
|
||
systematically misleading. RuView currently has no way to produce a MEASURED
|
||
vitals number: there is no reference-signal ingest, no time alignment between
|
||
CSI-derived estimates and a reference device, and no agreement statistics.
|
||
|
||
CLAUDE.md requires accuracy statements to be tagged MEASURED (with a
|
||
reproducer), CLAIMED, or SYNTHETIC. For vitals, MEASURED is currently
|
||
unreachable.
|
||
|
||
## Options considered
|
||
|
||
1. **Live BLE/ANT+ integration with reference devices.** Rejected for now:
|
||
drivers and pairing are a hardware/product concern; the blocking gap is
|
||
the evaluation math, not the radio link.
|
||
2. **File-based reference ingest + offline agreement analysis.** Chosen:
|
||
every consumer reference device (Polar, Garmin, oximeters) exports
|
||
timestamped series; a file boundary keeps the crate dependency-free and
|
||
the pipeline deterministic.
|
||
|
||
## Decision
|
||
|
||
Add a `groundtruth` module to `v2/crates/wifi-densepose-vitals`:
|
||
|
||
### 1. Reference series ingest
|
||
|
||
- `ReferenceSeries`: timestamped samples (unix millis + value) for one
|
||
measurand (`HeartRateBpm` or `BreathingRateBrpm`), with device metadata
|
||
(make/model, measurement principle). Parsed from CSV (`timestamp_ms,value`
|
||
with a header line); malformed rows are rejected with row-numbered errors —
|
||
untrusted file input validated at the boundary. Non-monotonic timestamps
|
||
are an error, not silently sorted.
|
||
|
||
### 2. Time alignment
|
||
|
||
- Constant-offset estimation by maximizing normalized cross-correlation of
|
||
the estimate series against the reference over a bounded lag window
|
||
(default ±30 s), on a common resampled grid (nearest-sample, no
|
||
interpolation of physiological values across gaps larger than a
|
||
configurable limit).
|
||
- Optional linear clock-drift fit (offset + rate) for long sessions.
|
||
Alignment parameters are reported, never silently applied.
|
||
|
||
### 3. Agreement metrics
|
||
|
||
- `AgreementReport`: n paired samples, coverage fraction (time where both
|
||
series had valid samples), MAE, RMSE, mean error (bias), Bland–Altman
|
||
95% limits of agreement, and percentage-within-tolerance (configurable,
|
||
default ±2 bpm HR / ±1 brpm breathing).
|
||
- Session scope is mandatory metadata: subject count, motion state
|
||
(static/moving), line-of-sight (LOS/NLOS/through-wall), distance band.
|
||
A report without scope cannot be constructed.
|
||
|
||
### 4. Evidence tagging
|
||
|
||
- `EvidenceGrade::Measured` is only constructible when the report carries a
|
||
reference device, non-zero paired samples, minimum coverage, and a
|
||
reproducer command string; otherwise the report grades as `Claimed` (real
|
||
data, no reference) or `Synthetic` (generated input). This mirrors
|
||
ADR-291's enforcement-in-types approach and the CLAUDE.md tagging rule.
|
||
|
||
## Consequences
|
||
|
||
- RuView can convert vitals claims from CLAIMED to MEASURED with a
|
||
reproducible offline analysis, session by session, scope by scope.
|
||
- Honest reporting will likely show heart-rate performance below marketing
|
||
intuition, especially NLOS/moving — that is the purpose.
|
||
- CSV ingest means a manual export step per session; acceptable at current
|
||
scale, and the format is the de-facto export of consumer reference gear.
|
||
- No clinical claim is implied: agreement statistics against consumer
|
||
reference devices are engineering evidence, not medical validation.
|
||
|
||
## Validation
|
||
|
||
- `cargo test -p wifi-densepose-vitals` — CSV rejection cases, alignment
|
||
recovery of known synthetic offsets/drifts, agreement metrics against
|
||
hand-computed fixtures, evidence-grade constructibility rules.
|
||
- `cargo bench -p wifi-densepose-vitals` — criterion benchmark for alignment
|
||
over hour-scale synthetic sessions.
|
||
- Real-session validation (ESP32 capture + chest strap) remains a follow-up
|
||
requiring hardware evidence per CLAUDE.md.
|